CONSTRUCTION CASE

FIX800 ETO (Ethylene Oxide) Leak Detector Installation — Hospital and Medical Sterilizer Rooms (Nationwide)

Nationwide · 2021.06.16

Applied Markets

Research Hospital

Installed Gases

ETO (Ethylene Oxide)

Project Details

FIX800 ETO (Ethylene Oxide) Leak Detector Installation — Hospital Sterilizer Rooms (Nationwide)

This article covers installation cases for the FIX800 ETO leak detector, widely used in hospital sterilizer rooms across the country.

What Is ETO (Ethylene Oxide)?

ETO (ethylene oxide), commonly referred to as "EO gas" or sterilization gas, is used primarily for sterilizing medical devices and instruments.

가스측정기 전문 완디코리아 WANDI®

ETO is highly effective as a sterilant, but even a small leak is acutely harmful. It is classified as a hazardous substance under Korea's OSHA, with regulated occupational exposure limits.

Hospital Sterilizer Room — Sterilizer and Gas Cylinder

가스측정기 전문 완디코리아 WANDI®

ETO sterilizer units with attached gas cylinders are found throughout hospitals. As the sterilizer operates using ETO from the cylinder, aging equipment or a loose valve can allow small amounts of gas to leak into the room.

가스측정기 전문 완디코리아 WANDI®

FIX800 Installation Process

ETO detector installation is straightforward. The unit is secured to the wall using a power drill, then power is connected.

가스측정기 전문 완디코리아 WANDI®

Once power is applied, the initial setup is essentially complete. Before finishing, correct power-on and zero calibration must be verified.

가스측정기 전문 완디코리아 WANDI®

After confirming proper operation, the facility manager is briefed on the rationale for the mounting location and the confirmed operational status before final finishing work begins.

가스측정기 전문 완디코리아 WANDI®

Once cable molding is complete, installation is finished. Total on-site time is typically around 30 minutes.

Ventilation Is Essential During Sterilizer Operation

ETO concentrations may rise during sterilizer operation cycles. Adequate ventilation in the sterilizer room is therefore mandatory — not optional.

Frequently Asked Questions

Q1. What are the occupational exposure limits for ETO?

Under Korea's OSHA: TWA 1 ppm and STEL 5 ppm. ETO is also classified as a carcinogen, meaning chronic low-level exposure carries long-term health risks even below the acute limits. Continuous monitoring in sterilizer rooms is strongly recommended.

Q2. Where should an ETO detector be installed?

ETO is approximately 1.5 times denser than air and tends to accumulate near floor level. Installation within 30 cm of the floor, near the sterilizer and gas cylinder valve connections, is generally recommended. Exact placement should be confirmed with a specialist based on the room layout.

Q3. Does the detector continue operating during sterilizer cycles?

Yes — continuous monitoring is the purpose. Small amounts of ETO may be detected during normal sterilizer operation, so alarm thresholds should be set appropriately to distinguish normal operation from a hazardous leak. Configuration should reflect the site's baseline ETO levels.

Q4. Is a detector alone sufficient without a ventilation system?

No. A detector identifies the presence of gas and triggers an alarm — it does not remove gas from the room. To protect personnel when an alarm occurs, ventilation must be available to exhaust the gas and enable safe evacuation. Detectors and ventilation systems are complementary and should always be deployed together.

Q5. Where else is ETO monitoring required beyond hospitals?

ETO is used in medical device manufacturing and sterilization facilities, food packaging sterilization plants, and spice and grain fumigation operations. All such facilities are required under Korea's OSHA to manage ETO exposure, making fixed detector installation the standard practice.


#ETODetector #EthyleneOxideMonitor #FIX800 #HospitalSterilizerSafety #MedicalDeviceSterilization #ToxicGasDetector #FixedGasDetector #IndustrialSafety #WANDI #Nationwide

Other Cases

Back to List